Test Device Slot; Usb Port; Operation; General - BD Veritor Plus Instructions D'utilisation

Masquer les pouces Voir aussi pour Veritor Plus:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 41
3.1.4

Test Device Slot

The test device slot is located on the right side of the BD Veritor Plus Analyzer (see arrow in figure below). Test devices are inserted by
aligning the insertion arrow on the device with the insertion arrow on the Analyzer. A distinctive "click" will be noted when the test device is
properly engaged in the Analyzer.
3.1.5

USB Port

The BD Veritor Plus Analyzer includes a USB On-The-Go port which is used for three applications:
Analyzer Update: the BD Veritor Plus Analyzer can be updated, using a USB flash drive that is provided by BD, to add capability to
read additional BD Veritor assays and to add new features. Instructions to perform these updates are provided with the USB flash
drive when such updates are available.
Printing: The optional printer is connected to the BD Veritor Plus Analyzers via this port. See section 2.4 for instructions.
Data Download: BD Veritor Plus Analyzers that are equipped with a BD Veritor InfoScan Module can be unlocked to enable secure
transfer of the test record to a computer via this port. See section 4.8 for instructions.
Data Transmission: BD Veritor Plus Analyzers that are equipped with the BD Veritor Plus Connect accessory can transmit test results
to a facility LIS over a secure connection via this port.
The BD Veritor Plus Analyzer can be externally powered only from the coaxial power port on the rear of the instrument.
4

Operation

4.1

General

Section 4.2 provides instructions for routine operation of the BD Veritor Plus Analyzer using the Analyze Now operational mode. In this
mode, the test device is prepared and developed outside of the Analyzer, then inserted to perform the final read. Section 4.3 provides
instructions for routine operation of the BD Veritor Plus Analyzer using the Walk Away operational mode. In this mode, the test device is
prepared outside the Analyzer but allowed to develop inside the Analyzer. The Analyzer initiates the final read at the appropriate time. Both
sections include instructions for operation with or without the BD Veritor InfoScan module installed. Sections 4.4 and 4.5 describe the
Quality Control (QC) test procedure in Analyze Now and Walk Away mode, respectively.
4.2
Using the BD Veritor Plus Analyzer – Analyze Now Operation
The BD Veritor Plus Analyzer is intended for use with BD Veritor Test Devices. Specific instructions for performing a BD Veritor Test can be
found in the package insert included with each test kit. After following those instructions the procedure for interpreting a test result with the
BD Veritor Plus Analyzer using Analyze Now operation is as follows:
1
Place the BD Veritor Plus Analyzer on a flat, dry, stable surface.
2
Ensure that the BD Veritor Plus Analyzer is not in direct sunlight or exposed to a bright light.
3
Check the BD Veritor Plus Analyzer for an inserted device. If a device is present, remove it from the Analyzer.
NOTE
The Analyzer cannot be powered through its USB port.
19