6
When the display window shows INSERT TEST DEVICE OR DOUBLE-CLICK BUTTON FOR WALK AWAY MODE, double-click the
power button.
7
If the BD Veritor Plus Analyzer has been configured with barcode ID – Enable Operator ID, it will display SCAN OPERATOR ID. When
this message is displayed, after verifying that the Operator ID to be scanned is correct, the operator should scan his/her identification
barcode. This message is displayed for up to 30 seconds, after which the process will restart with step 6. After a specific Operator
Identification has been recorded, that Operator Identification will be utilized for all subsequent tests until the BD Veritor Plus Analyzer
is powered off. On the next power cycle, the SCAN OPERATOR ID prompt will be repeated during the first test. A test result cannot be
generated if the Operator ID is not scanned when the Operator ID workflow option is enabled.
BD recommends reading Operator Identification barcodes with the BD Veritor Plus Analyzer positioned at the edge of a flat surface.
The barcode should then be moved toward the barcode window to be scanned. The scanned barcode value will be shown in the next
display window.
8
In the case of Quality Control (QC) Tests, the barcode on the printed label of the swab control supplied with the kit is used as
the Specimen ID.
The display window will now display SCAN SPECIMEN ID. When this message is displayed, the operator should scan the barcode on
the printed label on the swab control supplied with the kit. This message is displayed for up to 30 seconds, after which the process
restarts with step 6. A test result cannot be generated if the swab control barcode is not scanned.
BD recommends reading Quality Control (QC) barcodes with the BD Veritor Plus Analyzer positioned at the edge of a flat surface. The
barcode should then be moved toward the barcode window to be scanned. The scanned barcode value will be shown in the next
display window.
9
In the case of Quality Control (QC) Tests, the barcode on the printed label of the test device kit box must be scanned.
The display window will show SCAN KIT LOT NUMBER whether the kit information workflow configuration setting is enabled or
disabled for normal operation. When this message is displayed, the operator should scan the barcode on the exterior of the BD Veritor
Test Kit box. This message is displayed for up to 30 seconds, after which the process restarts with step 6. A test result cannot be
generated if the Kit Lot Number is not scanned.
If the BD Veritor Test Kit box label has two barcodes, scan only the upper barcode which starts with (17).
BD recommends reading Kit Lot Information barcodes with the BD Veritor Plus Analyzer positioned at the edge of a flat surface. The
barcode should then be moved toward the barcode window to be scanned. The scanned barcode value will be shown in the next
display window.
10
The display window will now show ADD CONTROL TO TEST DEVICE AND INSERT IMMEDIATELY. This message is displayed for up
to 3 minutes, after which the process restarts with step 6. Apply the prepared control to the test device sample well as instructed in the
test device kit package insert and immediately insert the test device into the slot on the right side of the Analyzer, aligning the insertion
arrow on the test device with the arrow above the slot. Insert the device fully until it stops. A distinctive "click" will be noted when the
device is properly aligned in the Analyzer. During this process the test device must remain horizontal to prevent spilling the specimen
out of the sample well.
11
The display window will now show DO NOT DISTURB TEST IN PROGRESS. The incubation time is determined based on the test
device barcode. The incubation time remaining is shown on the display. Do not disturb the test device or Analyzer during the
incubation period. Do not remove the test device; doing so will cause the test to abort.
12
After the incubation is complete, the BD Veritor Plus Analyzer will progress through two processing steps: a reading step, followed by
an analyzing step. The display will show the remaining time for each step as they are performed. Do not touch the Analyzer or remove
the test device during this time.
13
When the analysis is complete, the test result will be displayed with the control swab identification, the name of the test, a result, and
an overall "QC PASS" or "QC FAIL" indication. Refer to the package insert for the BD Veritor Test Device for specific result
interpretations. If a printer is connected to the unit and powered on, the test result will automatically be sent to the printer.
14
Verify that the test type and QC swab control are correct.
15
Once the test type and swab control are verified and the result noted, remove the test device by pulling it out. The display will show
INSERT TEST DEVICE OR DOUBLE-CLICK BUTTON FOR WALK AWAY MODE to indicate the Analyzer is ready to perform another
test.
16
When the BD Veritor Plus Connect accessory is installed, the ENVELOPE symbol will appear to indicate that results are being
transmitted over the secure connection. In the event that the BD Veritor Plus Analyzer cannot transmit the results to
BD Synapsys Informatics, it will queue all results to be transmitted and continuously attempt to transmit the results while it is powered
on. If the Analyzer is powered off while the ENVELOPE symbol is still present, the Analyzer will queue the result and transmit it the
next time that it is powered up. The symbol will disappear after the results have been transmitted.
17
To initiate a new test, repeat steps 6–16 above, or turn off the power by depressing the front panel power button for at least half a
second and releasing it. If the Analyzer is left unattended for 60 minutes, the BD Veritor Plus Analyzer will automatically shut off and
the test result will not be retained on the screen.
25