BD Veritor™ Plus Analyzer Instructions for Use
Result Number
34
none
none
none
none
none
none
34
Message
The BD Veritor Plus Analyzer
has detected that the QC test
procedure has not been
completed successfully,
because:
Test name and result
followed by:
QC FAIL
Defective USB flash drive or
UPDATE FILE
non-supported device
NOT FOUND
inserted into BD Veritor Plus
REMOVE USB DEVICE
Analyzer USB port.
INVALID
An invalid kit lot barcode was
BARCODE
detected by the BD Veritor
REINSERT
Plus Analyzer.
TEST DEVICE
INVALID
An invalid kit lot barcode was
KIT LOT
detected by the BD Veritor
BARCODE
Plus Analyzer.
OPERATOR ID
SCAN TIMEOUT
No barcode was scanned
REINSERT
within 30 seconds.
TEST DEVICE
SPECIMEN ID
SCAN TIMEOUT
No barcode was scanned
REINSERT
within 30 seconds.
TEST DEVICE
LOT NUMBER
SCAN TIMEOUT
No barcode was scanned
REINSERT
within 30 seconds.
TEST DEVICE
Possible Cause(s)
Ensure that the same test device type has been used
across all components (for example, a BD Veritor System
A mismatch was detected
Test Flu A+B test device must be used with the QC swab
between the QC swab
appropriate for the test device).
and test device (for
example, a Strep A QC
Ensure that the test device inserted was developed from
swab was run with an
the QC swab that was scanned with the barcode scanner.
RSV test device).
The test outcome did not
On first occurrence, repeat the QC procedure. On
match the expected
subsequent occurrences, contact BD.
outcome (for example, a
Strep A positive result
was expected but a Strep
A negative result was
detected).
If attempting to update the Analyzer, confirm that the USB
flash drive has been provided by BD.
Repeat the firmware update procedure as described in
the instructions included with the USB flash drive. On
subsequent occurrences, contact BD.
Ensure that the barcode scanned during this step is from
a valid BD Veritor Test Kit box label. For kits with two
barcodes, scan only the upper barcode that starts with
(17).
Ensure that the barcode scanned during this step is from
a valid BD Veritor Test Kit box label. For kits with two
barcodes, scan only the upper barcode that starts with
(17).
Reinsert test device and scan all barcodes within
30 seconds of the display prompt.
Reinsert test device and scan all barcodes within
30 seconds of the display prompt.
Reinsert test device and scan all barcodes within
30 seconds of the display prompt.
Corrective Action(s)