BD Veritor™ Plus Analyzer Instructions for Use
3.1.3

System Indicators

The meanings of on-screen displays used on the BD Veritor Plus Analyzer are as follows:
Symbol
Hourglass: An indication that the Analyzer is nearing end of life, either due to usage duration (time) or usage count.
In addition to the Hourglass icon, a descriptive message will appear on the screen after the power-on self-test is
complete. The message describes which end-of-life event is approaching and the remaining usage time or usage
count available. Each end-of-life event generates a unique message; insertion of a test device terminates all
messages and immediately initiates a test.
When this symbol is present, the user should contact their BD representative to order a new BD Veritor Plus
Analyzer.
The symbol switches from a continuous state to flashing when end of life is imminent.
See section 2.2 for an explanation of lifetime specifications.
Plug: When the Analyzer is properly connected to the facility power supply and powered on, the Plug icon will
appear on screen.
If the Plug icon is flashing, it indicates a failure in the battery charging circuit. Contact BD for assistance. The
Analyzer may continue to be used when connected to the facility power supply.
Battery: An indication of battery strength. The maximum battery strength is represented by four bars.
When the battery is nearly discharged, no bars are displayed and the battery outline flashes, indicating an imminent
need for recharge. While the battery is charging, the bars will cycle to indicate charging is underway. Once the
battery is fully charged, the cycling will stop and four bars will appear in a continuous state.
This symbol only appears on BD Veritor Plus Analyzers that are equipped with the BD Veritor Plus Connect, and will
appear when a BD Veritor Plus Analyzer is connected to a facility LIS.
This symbol only appears on BD Veritor Plus Analyzers that are equipped with the BD Veritor Plus Connect. The
envelope symbol appears when test results are waiting to be transmitted. A flashing envelope indicates that data are
in the process of being transmitted.
18
Meaning