If the BD Veritor Plus Connect accessory is installed:
The ENVELOPE symbol will appear to indicate that results are being transmitted. In the event that the BD Veritor Plus Analyzer cannot
transmit the results to BD Synapsys Informatics, it will queue all results to be transmitted and continuously attempt to transmit the
results while it is powered on. If the Analyzer is powered off while the ENVELOPE symbol is still present, the Analyzer will queue the
result and transmit it the next time that it is powered up. The symbol will disappear after the results have been transmitted.
10
To initiate a new test, repeat steps 5–9 above, or turn off the power by depressing the front panel power button for at least half a
second and releasing it. If the Analyzer is left unattended for 15 minutes (when operating on the internal battery), or 60 minutes (when
operating on the external power adapter), the BD Veritor Plus Analyzer will automatically shut off and the test result will not be retained
on the screen.
4.3
Using the BD Veritor Plus Analyzer – Walk Away Operation
The AC power adapter must be connected to the Analyzer and plugged into a facility power source to use
If the Analyzer is left unattended for 60 minutes after completion of the test run, the power will automatically shut off and
The BD Veritor Plus Analyzer is intended for use with BD Veritor Test Devices. Specific instructions for performing a BD Veritor Test can be
found in the package insert included with each test kit. After following those instructions up to sample preparation in the reagent tube, the
procedure for interpreting a test result with the BD Veritor Plus Analyzer using Walk Away operation is as follows. DO NOT DISPENSE THE
SPECIMEN INTO THE TEST DEVICE SAMPLE WELL AT THIS TIME.
1
Place the BD Veritor Plus Analyzer on a flat, dry, stable surface.
2
Ensure that the BD Veritor Plus Analyzer is not in direct sunlight or exposed to a bright light.
3
Ensure that the AC power adapter is connected to the Analyzer and plugged into a facility power source.
4
Check the BD Veritor Plus Analyzer for an inserted device. If a device is present, remove it from the Analyzer.
5
Press the front panel power button. The BD Veritor Plus Analyzer will complete a self-test before it is ready for use. After the self-test
completes and any temporary messages are presented, the display window shows INSERT TEST DEVICE OR DOUBLE-CLICK
BUTTON FOR WALK AWAY MODE.
If an InfoScan is installed:
The display window shows SCAN CONFIG BARCODE for 2 seconds after the self-test. This is an optional step, no action is required.
To perform instrument configuration, refer to section 4.6 for configuration instructions.
6
When the display window shows INSERT TEST DEVICE OR DOUBLE-CLICK BUTTON FOR WALK AWAY MODE, double-click the
power button.
If an InfoScan is installed:
a
If the BD Veritor Plus Analyzer has been configured with barcode ID – Enable Operator ID, it will display SCAN OPERATOR ID.
When this message is displayed, after verifying that the Operator ID to be scanned is correct, the operator should scan his/her
identification barcode. This message is displayed for up to 30 seconds, after which the process will restart with step 6. After a
specific Operator Identification has been recorded, that Operator Identification will be utilized for all subsequent tests until the
BD Veritor Plus Analyzer is powered off. On the next power cycle, the SCAN OPERATOR ID prompt will be repeated during the
first test. A test result cannot be generated if the Operator ID is not scanned when the Operator ID workflow option is enabled.
BD recommends reading Operator Identification barcodes with the Analyzer positioned at the edge of a flat surface. The barcode
should then be moved toward the barcode window to be scanned. The scanned barcode value will be shown in the next display
window.
NOTE
Walk Away operation.
NOTE
the test result will not be retained on the screen.