BD Veritor™ Plus Analyzer Instructions for Use
Result Number
15
16
17
18
19
20
21 / 22
23
32
Message
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 15
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 16
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 17
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 18
BACKGROUND INVALID
The assay background area
CONSULT MANUAL
is outside of the normal
ERROR 19
range.
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 20
The shelf life of the
BD Veritor Plus Analyzer has
expired. The maximum shelf-
life of the BD Veritor Plus
Analyzer from the date of
manufacture is 34 months
ANALYZER
(regardless of time of first
EXPIRED
use).
NO DAYS
REMAIN
The operational life of the
BD Veritor Plus Analyzer has
expired. The maximum
number of months from first
use is 24 months.
The operational life of the
ANALYZER
BD Veritor Plus Analyzer has
EXPIRED
expired. The maximum
NO TESTS
number of tests performed by
REMAIN
a BD Veritor Plus Analyzer is
3,500 runs.
Possible Cause(s)
Contact BD.
Contact BD.
Contact BD.
Contact BD.
Remove the device and cycle the instrument power.
If the instrument displays an error during power-on,
contact BD.
If no error message appears, obtain an additional
sample and repeat the assay If the issue recurs,
contact BD.
Contact BD.
Contact your distributor to order a new BD Veritor Plus
Analyzer.
or
Contact your distributor to order a new BD Veritor Plus
Analyzer.