Controls; Control Panel; Apparatus - Enraf Nonius Radarmed 650+ Mode D'emploi

Masquer les pouces Voir aussi pour Radarmed 650+:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

6.1. Control panel

(See jacket flap)
[1] Indication LED therapy mode
The illuminated LED indicates the selected mode;
continuous or pulsed.
[2] Selector therapy mode: continuous
or pulsed microwave
The unit automatically switches to continuous mode
when switched on or at the end of treatment. It is
possible to switch from continuous to pulsed mode
and visa versa during treatment.
[3] Display treatment time
The display shows the set c.q. remaining treatment
time in minutes. When the time is expired a zero
appears on the display, a sound signal is audible
and the HF-power switches off. A flashing dot in the
display means the clock is running and treatment
started.
[4] Treatment time setting selectors
These selectors are used to set the treatment time.
[5] Power display
The display shows the set power. A flashing zero on
the display indicates that no treatment time has been
set. In this case it is not possible to set the power.
[6] Power setting selectors
These selectors are used to set the power. It is only
possible to set the power if also a treatment time
has been set.
[7] Stop button
Button to end a treatment premature. After opera-
tion of this button the treatment time runs down to
zero, a sound signal is audible and the HF-power is
shut off.

6.2. Apparatus

[8] On/off switch
With this switch the Radarmed is turned on/off.
[9] Ventilation slots
These openings must be kept clear for ventilation.
[10] Connection points electrode arm
The electrode arm can be connected at choice at
the left or right side of the unit (see paragraph 7.6 of
this manual).
[11] Hook for mains cable
Hook to hang up the mains cable when the unit is
out of operation.
[12] HF-connection
Connection for the HF-cable (see paragraph 7.8 of
this manual).
[13] Connection mains cable
Connect the Radarmed to a grounded safety socket
of a voltage corresponding to the data mentioned
on the type plate of the unit.
[14] Connection for potential equalization cable
A special cable can be supplied for areas where lo-
cal regulations require connection of a potential
equalization cable.
[15] Type plate
The type plate gives details of the model, series-
and serial number (for service, warranty etc.) as well
as ratings such as mains voltage and power con-
sumption.