7.7. Connection radiator

To attach or remove the radiator, push the black lock-
ing knob (on top the radiator arm). When fitting the
radiator, the locking knob must click into place.

7.8. Connection HF-cable

7.8.1. Connection to the unit
The HF-cable is connected through connection point
[12] to the unit by means of a screw thread connec-
tion.
The length of the HF-cable is critical. Therefore only
use the by Enraf-Nonius supplied cable.
7.8.2. Connection to the radiator
The HF-cable is connected to the radiator by means
of a "fast" coupling.
To plug on the connector
Push on connector
Check to ensure that connector is secure
To unplug the connector
Pull back ring
Pull out connector

7.9. Switching on and self test

• Switch on the unit using the on/off switch [8].
• Immediately after switching on, the unit carries
out a self test
Check that:
• all LED's light up briefly;
• at the end of the test a beep can be heard.
Contact your supplier if this is not the case.