Mettler Toledo IND570 Série Guide D'installation page 21

Masquer les pouces Voir aussi pour IND570 Série:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Note that, in addition to a 25mm gland, there are three 16mm glands with a smaller cable entry
hole and two with a larger cable entry hole. Make sure to choose the gland with the correct hole
size based on the cable used. To maintain the IP rating of the enclosure, the gland must tighten
securely around the cable. Table 2-3 lists the range of cable sizes that fit each gland.
For proper installation of a panel-mount IND570xx with an IDNet or SICSpro scale interface inside
an enclosure, the cable must enter the enclosure through an ATEX/IECEx certified gland. The gland
must be large enough to allow the IDNet or SICSpro connector from the base to pass through the
center and then seal tightly around the cable. This ATEX/IECEx gland is not provided by METTLER
TOLEDO.
The standard IND570 terminal offers options for external connectors to extend the USB port and the
Ethernet port to the outside of the enclosure. These options MUST NOT be used in the IND570xx
installation when the terminal will be located inside the Division 2 or Zone 2/22 area.
2-4
METTLER TOLEDO IND570xx Division 2 and Zone 22 Installation Manual
COM2/COM3
/DIO
DIO/PLC
Table 2-2 : Gland Torque Specifications
Gland Size
Hole Size
M16 – smaller
M16 – larger
Figure 2-4: Gland Locations
Torque
M16
5.5 Nm (49 lbf-in)
M25
12.0 Nm (8.9 lbf-ft)
Table 2-3 : Gland Cable Hole Size
Cable Diameter
3 - 7 mm
6 - 10 mm
M25
3 – 4 mm
Not Used
Ethernet
COM1/ PLC
PLC
30205321
|
05
|
03/2019

Publicité

Table des Matières
loading

Table des Matières