Recommended Separation Distances Between Portable And Mobile Rf Communications And Adherus Autospray Et Dural Sealant - HyperBranch Adherus AutoSpray ET Dural Sealant NUS-009 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 64
Recommended Separation Distances Between Portable and Mobile RF
Communications and Adherus AutoSpray ET Dural Sealant
Adherus AutoSpray ET Dural Sealant is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or user of the Adherus AutoSpray ET Dural Sealant device can help
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Adherus AutoSpray ET Dural Sealant device as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2√P
0.12
0.38
1.2
3.8
12
m
80 MHz to
800 MHz
d = 1.2√P
0.12
0.38
1.2
3.8
12
15
800 MHz to
2.5 GHz
d = 2.3√P
0.23
0.73
2.3
7.3
23