7. EMC / Technical description (cont.)
Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 2)
Electromagnetic immunity
The electric breastpump Swing is intended for use in the electromagnetic environment
specified below. The customer or the user of the electric breastpump Swing should assure
that it is used in such an environment.
Immunity test
Electrostatic
discharge ( ESD )
IEC 61000-4-2
Electrical fast
transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
No input/output lines
a
No earth, class II equipment
b
32
IEC 60601
test level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input /
output lines
± 1 kV line( s ) to
line(s)
± 2 kV line( s ) to
earth
Electromagnetic
Compliance
environment
level
– guidance
Floors should be wood,
concrete or ceramic tile.
± 6 kV contact
If floors are covered with
± 8 kV air
synthetic material, the
relative humidity should
be at least 30%.
Mains power quality
± 2 kV for power
should be that of a
supply lines
typical commercial or
Not applicable
a
hospital environment.
± 1 kV
Mains power quality
differential
should be that of a
mode
typical commercial or
Not applicable
b
hospital environment.