Troubleshooting

Symptom
"Beeping" sound when
using the bed
One or more bed
functions inoperative
All functions inoperative
except for raising
mattress platform
Bed is difficult to
manoeuvre around
corners
All indicators on ACP lit
or flashing
If the equipment fails to operate correctly, the table
below suggests some simple checks and corrective
actions. If these steps fail to resolve the problem,
contact ArjoHuntleigh or an approved service agent.
Possible cause
Bed is operating from
the backup battery
Function(s) locked on
ACP
Height control
software error
Brake pedals in "steer"
position
Duty cycle of electrical
system exceeded
Action
Check the power supply
cord is plugged in and
the electricity supply is
OK
Check fuse in mains
plug (where fitted)
Unlock function(s) on
ACP
Raise the mattress
platform to maximum
height to reset software
Place brake pedals in
the "free" position
Refer to section Duty
cycle lockout in
Chapter 4
51