Indications; Contra-Indications - Arjohuntleigh Enterprise 5000X Notice D'utilisation

Masquer les pouces Voir aussi pour Enterprise 5000X:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Contra-

indications

WARNING
If the combined weight of the mattress and accessories exceeds
65kg, the maximum patient weight must be reduced accordingly.
The safe working load is calculated as follows (in
accordance with EN 60601-2-52:2010):
Maximum patient weight
Mattress
Accessories (including attached loads)
TOTAL
At the discretion of the carer, patients taller than 190cm
may be accommodated by extending the bed - refer to
"Bed length adjustment" in Chapter 4. Ensure that the
patient's height does not exceed the "In-bed length"
shown in Chapter 7.
The safe working load of the bed is
250kg.
The maximum recommended
patient weight is 185kg.
The recommended patient height
is between 146cm and 190cm.
185kg
20kg
45kg
250kg