Troubleshooting - Arjohuntleigh ENTERPRISE E8000X Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Troubleshooting

If the equipment fails to operate correctly, the table below
suggests some simple checks and corrective actions. If these
steps fail to resolve the problem, contact ArjoHuntleigh or an
approved service agent.
Symptom
"Beeping" sound when
using the bed
One or more bed
functions inoperative
Bed is difficult to
manoeuvre
All indicators on ACP lit
or flashing
Mattress platform
cannot be lowered
All functions remain
locked after connecting
mains power after a
near flat battery (ACP
battery indicator was red
before mains power was
connected)
Bed movement function
buttons do not respond
46
Possible Cause
Bed is operating from
the backup battery
Function(s) locked on
ACP
Brake pedals in "steer"
position
Duty cycle of electrical
system exceeded
Height control software
error
Function(s) locked on all
controls due to a low
power state
Control software error
Action
Check the power supply
cord is plugged in and
the electricity supply is
OK
Check fuse in mains
plug (where fitted)
Unlock function(s) on
ACP
Place brake pedals in
the "free" position
Refer to the section
Duty cycle lockout on
page 39
Raise the mattress
platform to maximum
height to reset software
To unlock all functions
connect mains power
and then press the
Function Lock button
twice in quick
succession then select
the function(s) to unlock
Disconnect and then
reconnect the mains
power to clear bed
software errors