Troubleshooting And Alarm Conditions - Arjohuntleigh Alpha Active 3 Manuel D'utilisation

Masquer les pouces Voir aussi pour Alpha Active 3:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

8. Troubleshooting and Alarm Conditions

The following table provides a troubleshooting and alarm condition guide for the
system in the event of malfunction. These alarms do not cause any
Alpha Active 3
delay or interruption in therapy.
Indicator
LOW PRESSURE
POWER FAILURE
SERVICE
Possible Cause
The tube-set is not
connected properly.
CPR not fully closed.
There is a leak in the
system.
Power has been
removed from the
pump.
Pump has detected an
internal fault, such as
the gearbox failure.
16
Remedy
Check the tube-set
connector and ensure
it is securely fitted to
the pump.
Close CPR unit.
Call service engineer.
Re-apply power or
switch the pump off.
Switch the pump off
and call service
engineer.
Priority
Low priority
According to
IEC60601-1-8
Low priority
According to
IEC60601-1-8
Low priority
According to
IEC60601-1-8

Publicité

Table des Matières
loading

Table des Matières