Electromagnetic Compatibility; Guidance And Manufacturer's Declaration - Electromagnetic Emissions - Arjohuntleigh Flowtron ACS900 Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Electromagnetic Compatibility

Guidance and manufacturer's declaration - electromagnetic emissions

The pump is intended for use in the electromagnetic environment specified below. The customer or the
user of the pump should assure that it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Recommended separation distances between portable
and mobile RF communications equipment and the pump
The pump is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the pump can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the pump as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Compliance
Group 1
Class A
Class A
Complies
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2√P
0.12
0.38
1.2
2.8
12
30
Electromagnetic environment - guidance
The pump uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
The pump is suitable for use in all establishments
other than domestic and those directly connected
to the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
m
80 MHz to 800 MHz
d = 1.2√P
0.12
0.38
1.2
3.8
12
800 MHz to 2.5 GHz
d = 2.3√P
0.23
0.73
2.3
7.3
23

Publicité

Table des Matières
loading

Table des Matières