It is recommended that the following checks are carried
out at the start of, and throughout, the therapy:
• Check the LCD screen icons to confirm that the
correct type of garment(s) have been connected.
• During garment inflation, check the LCD screen to
confirm that there are no fault indicators displayed
and that the correct pressure is being supplied. The
default target inflation pressures for the different
garments are:
• Foot garment: 130 mmHg.
• Uniform calf and calf & thigh garments:
40 mmHg.
• Sequential calf and calf & thigh garments:
45 mmHg.
• Check that there are no kinks in the tubeset.
• Check that the tubeset and connectors do not cause
the patient any discomfort.
• Regularly check that the garments remain correctly
fitted to the patient.
The pump will inflate and deflate each garment in turn,
as follows, starting with the garment connected to
tubeset connector 1. The garment pressure is displayed
in the top left side of the screen. The Patient Run Hours
are displayed in the bottom right of the screen.
The following example shows two foot garments
fitted to the pump.
1. Both garments are initially deflated, and the
garment icons are shown as outlines.
2. Foot garment 1 is inflated to the target pressure,
with an inflate time of 3 seconds and a deflate time
of 27 seconds. The foot garment icon is black while
it is inflated.
14