Audible Alarm Status
Audible Alarm Muted
Warning and Alarm
Indications
LED Indicators on the
Pump Case
LED Colours
Extinguished
Green
Yellow
When an alarm which is able to be muted is shown on
the LCD screen, this icon is shown to indicate that the
alarm can now be muted.
When the pump is in Standby, this icon indicates that the
alarm volume can be changed (refer to "To Change the
Audible Alarm Volume Setting" on page 17).
This icon indicates that the audible alarm has been
muted.
The following five icons indicate that a fault has been
detected in the system. Refer to "Troubleshooting" on
page 21) for a list of fault conditions, the corresponding
warning and alarm indications, and possible corrective
actions to rectify the faults.
There are additional LED indicators on the pump to
show pump status and alarms:
• Two sets on the front of the pump.
• One set on the underside of the pump.
Their status is as follows:
Pump Status
Off or Standby
• No fault detected.
Run
• Warning only detected.
Run
Fault detected with full audible and visual alarm.
Alarm detected which may cause an
interruption in therapy.
Low pressure or garment leak.
Kinked or blocked tubeset.
High temperature.
Hardware fault detected or periodic service
required.
Warnings/Alarms
- -