Faulty Tubeset
High Temperature
Description and Corrective Action
The Faulty Tubeset warning and alarm are activated
if the pump detects a fault in the tubeset.
For both conditions:
• Switch off the pump.
• Call the service engineer.
If the pump is in Standby when the faulty tubeset is
detected:
• A warning is activated.
The leg icon and garment icon (if present) will
flash, and show a faulty tubeset and affected
connector.
Therapy cannot be started.
If the pump is in the Run state when the faulty
tubeset is detected:
• An alarm is activated.
The leg icon and garment icon (if present) flash
and show a faulty tubeset and affected
connector.
The LED indicators on the front and underside of
the pump change to yellow.
Therapy can continue as if the same garment(s)
is attached.
• When the operator next stops the pump and puts
it into Standby, the alarm reverts back to the
Standby warning screen above with the flashing
leg icon and garment icon (if present).
Therapy cannot be restarted.
This warning occurs if the temperature inside the
pump exceeds 55°C. The High Temperature icon is
displayed and an audible alarm sounds. Therapy
can continue.
Make sure the pump is not close to a heat source, or
covered by blankets.
If the temperature inside the pump exceeds 60°C,
this screen is shown. Therapy is stopped.
Switch off the pump and allow pump to cool. Then
switch back on and continue therapy.
If either alarm continues, switch off the pump
and call the service engineer.