Owner's Guide
VII. Order Information
Unit:
Catalog No .
Voltage
60034537
120V
Inserts: Frequency 25 kHz
Catalog No .
Description
60034383
#10 Universal
60034373
#10 SuperSoft
60034379
#10 OptiMist
60034385
#10 Universal Slim
60034377
#10 Slim SuperSoft
60034387
#10 Slim OptiMist
60034375
#10 Slim SuperSoft (Optimist)
60034381
#1000 Universal Triple Bend
Inserts: Frequency 30 kHz
Catalog No .
Description
60034384
#10 Universal
60034374
#10 SuperSoft
60034380
#10 OptiMist
60034386
#10 Universal Slim
60034378
#10 Slim SuperSoft
60034388
#10 Slim OptiMist
60034376
#10 Slim SuperSoft (Optimist)
60034382
#1000 Universal Triple Bend
Replacement Parts:
60036143
Replacement O-Ring Gaskets
60036144
Water Filter Replacement Kit
60034538
US200 Replacement Core O-Ring
60034539
US200 Handpiece Sheath
60034540
US200 External Power Supply
60034542
US200 Footswitch
60034543
US200 Handpiece Assy
60036145
US200 Water Line
VIII. Troubleshooting
Generally, check all lines and connections to and from the unit .
Check the settings on the unit's knobs . Although service and repair
of the BioSonic US200 ultrasonic scaler should be performed by
authorized personnel, the following are some basic troubleshooting
procedures that may help avoid unnecessary service calls .
Frequency
Power
50/60 Hz
65W max
A. Unit will not operate:
LED Status
Off
Qty
1
1
1
1
1
1
1
1
Qty
1
Green
1
1
1
1
1
1
1
Blinking Green
and Orange
12
10
3
1
1
1
1
1
B. Unit will operate but LED is not solid green or orange:
LED Status
Blinking Green
(in Normal
Mode) or
Blinking Orange
(in Turbo Mode)
Potential Solution
Ensure the power cord is fully seated in the
receptacle on the power adapter .
Ensure the power adapter cord is fully seated
in the DC receptacle on the back of the unit .
Ensure the power cord is plugged into an
active and appropriate AC receptacle .
Ensure the power knob is in the ON position by
checking that the power knob is not rotated
fully counterclockwise . Knob should "click"
when rotated clockwise, turning power on .
Ensure the power adapter being used is the
one shipped with the unit .
Release the footswitch and reengage . The unit
will not activate if the footswitch is depressed
when the unit is turned on .
Ensure the handpiece connector is fully seated
in the receptacle on the front of the unit . A
"click" will be felt when it is properly inserted .
Ensure the footswitch is fully inserted in the
receptacle in the back of the unit . A "click" will
be felt when it is properly inserted . Inspect the
footswitch and cord for any obvious damage .
Ensure an appropriate 25 kHz or 30 kHz insert
is fully seated in the handpiece .
Ensure adequate water flow to the handpiece .
Replace the insert . It may be worn, clogged, or
damaged .
Potential Solution
Ensure adequate water flow to the
handpiece . Ensure water supply to the back of
the unit is properly connected and not kinked
or pinched . Ensure water filter is not clogged .
If clogged, replace with new filter . Adjust water
control knob on front of unit until an appropri-
ate mist is observed .
Ensure an appropriate 25 kHz or 30 kHz insert
is fully seated in the handpiece .
Reduce power setting .
Replace the insert . It may be worn, clogged, or
damaged .
11