5.5 Temperature control and mixing
After switching on, the values used the last time the unit was used before switching off will
be shown on the display. The actual temperature flashes. The unit is now in basic temperature
control mode.
The Mixer HC can cool to 13 °C below room temperature.
5.5.1
Basic temperature control
Set the temperature with the "TEMP" arrow keys or press "START/STOP", if you do not
i
want to change the nominal temperature.
The device starts the temperature control. On the display, the nominal temperature is
displayed on the left, the actual temperature on the right. The temperature between 1°C
and 99°C is then adjusted immediately. When the nominal value is reached, only this value
will be displayed.
The temperature can be adjusted gradually by pressing the key briefly, or adjusted continu-
ously by pressing and holding the key.
You can also change the nominal temperature while the unit is running. The run will be
continued with the re-set nominal temperature.
5.5.2
Operation without temperature control
Press the "TEMP" arrow key until off appears on the display.
i
The unit remains at room temperature and does not heat. You can now use the unit as a
mixer without temperature control.
5.5.3
Mixing
The mixing frequency (-- rpm) is displayed on the right of the display next to the temperature.
1.
Set the mixing frequency with the „MIX" arrow keys.
You can set the mixing frequency to 0 or between 300 and 1500 rpm. Pressing the key
once will adjust the frequency by 50 rpm.
Depending on the software version, the unit will automatically recognise the mounted
exchangeable thermoblock. The mixing frequency specified for the exchangeable ther-
moblock cannot be increased in this case.
20
· Mixer HC · User Manual