6.2 Error messages
Symptom / message
MOTOR ERR
SPEED ERR
TOO HOT!
T-SENSOR ERR
TEMP ERR!
EEChksumERR
RAM ERR
EEprom ERR
HEAVY LOAD
24
· Mixer HC · User Manual
Cause
Motor is not turning.
Motor is turning at over-
speed.
Thermoblock hotter than
110 °C.
Temperature sensor error.
Nominal temperature is not
reached.
Memory error.
The motor does not reach
nominal speed.
Remedy
1.
Switch unit off and wait
for 10 seconds.
2.
Switch the device on.
1.
Switch unit off and wait
for 10 seconds.
2.
Switch the device on.
1.
Switch unit off and wait
for 10 seconds.
2.
Switch the device on.
1.
Switch unit off and wait
for 10 seconds.
2.
Switch the device on.
1.
Switch unit off and wait
for 10 seconds.
2.
Switch the device on.
1.
Switch unit off and wait
for 10 seconds.
2.
Switch the device on.
Check if the movement of the
i
exchangeable thermoblock
is obstructed (eg. by jammed
object).