2.
Press the "START/STOP" key to start the mixing process.
During the run, the green indicator lamp in the "START/STOP" key lights up.
3.
To finish the mixing process, press "START/STOP" again.
After setting the required parameters, the unit will start mixing and/or the temperature
control after pressing the "START/STOP" key. The temperature control may have also
started by re-setting the nominal value for the temperature control.
When the pre-set time has elapsed or the "START/STOP" key is pressed again, mixing is
completed. The temperature is, however, kept constant.
5.5.4
Short Mix
Keep the "SHORT MIX" key pressed
i
The unit mixes with the set frequency until the "SHORT MIX" key is released. During the
first minute, time counting starts in second intervals and then changes to minute intervals.
5.5.5
Interrupting operation
1.
Keep the "START/STOP" key pressed for more than two seconds.
Mixing and time counting are interrupted. You can now carry out manual operations (eg.
adding reagents, replacing tubes).
2.
To continue the run, press the "START/STOP" key again.
5.5.6
Interval Mix
The "INTERVAL MIX" function enables a continuous alternation of mixing activity and mixing
pause. The device carries out the mixing procedure, interrupted by pauses, until the set total
time has elapsed. The times for mixing and the mixing pause (from 3 seconds to 99 minutes 59
seconds) can be set independently
1.
Press the "INTERVAL MIX" key for two seconds or more.
The indicator lamp of the key lights up in red. On the display the
interval function is now activated.
2.
Using the "TIME" arrow keys, set the total time for the interval function.
3.
Briefly press the "INTERVAL MIX".
The
sign flashes on the display.
4.
Using the "TIME" arrow keys, set the mixing phase.
sign appears. The
Mixer HC · User Manual ·
21