Measuring procedure
1. Place the chest piece 6 underneath the cuff 2 or 1-2 cm
below it. Make sure the chest piece is in contact with the skin
and lies on the brachial artery.
2. Plug in the ear piece 7 and check, if the chest piece is placed
correctly, so that the Korotkoff sounds appear the loudest.
3. Close the valve 4 on the pump ball 3 by turning the screw
clockwise. Do not over-tighten.
4. Take the pump ball 3 in your free hand (the arm you are not
using to measure) and pump up the cuff. Watch the pressure
indicator on the manometer 1 and pump up to approx.
40 mmHg higher than the expected systolic value (the upper
value).
Inflate to 200 mmHg if you are not sure about the expected
value.
5. Open the valve 4 slowly by turning the screw counter clock-
wise whilst holding the stethoscope chest piece 6 on the
brachial artery. Listen carefully as the cuff begins to deflate.
Note the reading on the manometer 1 as soon as you hear a
faint, rhythmic tapping or thumping sounds. This is the
systolic blood pressure reading.
6. Allow the pressure to continue dropping at the same deflation
rate. Note the reading on the manometer 1 as soon as the
thumping sound stops. This is the diastolic blood pressure
reading.
7. Deflate the cuff completely.
8. Repeat the measurement at least two further times and record
your values, date and time immediately after finishing the
measurements.
9. Remove the cuff and the stethoscope.
4. Malfunction / Troubleshooting
If problems occur when using the device, the following points
should be checked and if necessary, the corresponding measures
are to be taken:
Microlife BP AG1-30
Description
Potential cause and remedy
The sound transmis-
Check the ear pieces if they are dirty
sion is poor, distorted
or there is extraneous
noise.
Check the tube if it is broken or
Check the chest piece if there is any
Make sure the chest piece is in
The pressure does not
Make sure that the valve is closed.
rise although the pump
Make sure the cuff is properly
ball 3 is pumping.
Check if the cuff, tube and/or bulb is
The deflation rate can
Disassemble the valve from pump ball
not be set to 2-
to check if there is any blockage in the
3 mmHg/sec. by
airway of the valve. Clean the blockage
adjusting the valve 4.
and try again. If it still does not work,
replace it to avoid inaccurate readings.
The manometer
Make sure that the valve is
needle is not at 0 ± 3
mmHg at rest.
If still more than 3 mmHg deviation,
If you think the results are unusual, please read through the
information in «Section 1.» carefully.
5. Safety, care, accuracy test and disposal
Safety and protection
Follow instructions for use. This document provides important
product operation and safety information regarding this device.
Please read this document thoroughly before using the device
and keep for future reference.
or cracked. If not, make sure you
wear them properly.
twisted.
damage.
contact with the skin and lies on the
brachial artery. Clean or replace any
defective parts if found to avoid inac-
curate reading.
connected to bulb and manometer.
leaking. Replace the defective parts
if any.
completely open for zero check.
contact your dealer to recalibrate the
manometer.
EN
3