1
2
Fig. 19: Mains input fuse
1
Fuse holder
2
Snap lock
Model
MIKRO 200
MIKRO 200
MIKRO 200 R
MIKRO 200 R
10
Disposal
10.1
General instructions
AB2400en
The mains fuses are located next to the mains switch.
The mains switch is in switch position
1.
Disconnect the mains cable from the device plug.
2.
Press the snap lock (
3.
Replace the defective mains input fuses.
Only use fuses with the nominal value specified for the type: see the
table below.
4.
Push in the fuse holder (
5.
Reconnect the device to the mains.
Type
2400
2400-01
2405, 2405-07
2405-01
The device can be disposed of via the manufacturer.
A Return Material Authorisation (RMA) form must always be
requested for a return.
If necessary, contact the Technical Service Department of
the manufacturer:
−
−
−
−
−
Risk of pollution and contamination for people and the envi-
ronment.
When disposing of the centrifuge, people and the environ-
ment may be polluted or contaminated by incorrect or
improper disposal.
−
The device is intended for the commercial sector ("Business to Business" -
B2B).
According to Directive 2012/19/EU, the devices may no longer be disposed
of with household waste.
The appliances are assigned to the following groups according to the Stif-
tung Elektro-Altgeräte Register (EAR (German foundation under civil law)):
Rev.: 13 / 01.2024
[O]
2 ) against the fuse holder ( 1 ) and pull it out.
1 ) until the snap lock engages.
Fuse
T 3.15 AH/250 V
T 5 AH/250 V
T 5 AH/250 V
T 8 AH/250 V
Andreas Hettich GmbH & Co. KG
Föhrenstraße 12
78532 Tuttlingen, Germany
Phone: +49 7461 705 1400
E-Mail: service@hettichlab.com
WARNING
Removal and disposal may be carried out only by a
trained and authorized service personnel.
Disposal
Order no.
E997
E914
E914
E738