7.3.3
Precooling the rotor
Starting
Set
7.4
Machine Menu
7.4.1
Querying system information
Parameter query
AB2400en
1.
Press and hold the
't/min = X ' is displayed after 8 seconds.
[Rotary knob] to set the delay time.
2.
Use the
[START/IMPULS] button.
3.
Press the
The setting is stored.
'*** ok ***' is displayed briefly.
[STOP/OPEN] button twice or wait 8 seconds to exit the
4.
Press the
menu.
The rotor is stationary.
[Cooling] button.
1.
Press the
[STOP/OPEN] button.
2.
Press the
Precooling of the rotor is terminated.
Ramp-down takes place with the selected brake level.
The brake level is displayed.
The precooling speed is adjustable from 500 RPM up to the maximum rotor
speed in increments of 10 RPM. It is preset to 10000 RPM.
■
The rotor is stationary.
■
The lid is open.
1.
Press and hold the
't/min = X ' is displayed after 8 seconds.
[Cooling] button.
2.
Press the
Precooling speed
[Rotary knob] to set the precooling speed.
3.
Use the
[START/IMPULS] button.
4.
Press the
The setting is stored.
'*** ok ***' is displayed briefly.
[STOP/OPEN] button twice or wait 8 seconds to exit the
5.
Press the
menu.
The rotor is stationary.
1.
Press and hold the
'SOUND/BELL' is displayed.
[SELECT] button repeatedly until 'FU/CCI – S.' is displayed.
2.
Press the
Program version for the frequency converter
[SELECT] button repeatedly until 'HOURS' is displayed.
3.
Press the
Internal operating hours (the time during which the centrifuge was
switched on)
4.
Turn to the right with the
'STARTS' is displayed.
Number of centrifugation runs
Rev.: 13 / 01.2024
[Cooling] button.
[Cooling] button.
'RPM = XXXX' is displayed.
[SELECT] button for 8 seconds.
[Rotary knob] .
Software operation
29 / 39