Software operation
6.5.3
Short-time centrifugation
6.6
Quick stop function
7
Software operation
7.1
Centrifugation parameters
7.1.1
Relative centrifugal force, RCF
26 / 39
3.
Ramp-down takes place with the selected brake level after the time
has elapsed or if the centrifugation run is cancelled.
The brake level is displayed.
An audible signal sounds when the rotor comes to a standstill.
'OPEN' 'OEFFNEN' is displayed.
The right side of the
centrifuge is in ramp-down.
The left side of the
at a standstill.
The light on the
[STOP/OPEN] button go out.
Personnel:
■
Trained user
1.
Press and hold the
[START/IMPULS] button lights up during the centrifugation
The
run.
Timing starts at 00:00.
The rotor speed or the RCF value, the temperature in the centri-
fuging chamber (only for centrifuges with cooling) and the elapsed
time are displayed during the centrifugation run.
[START/IMPULSE] button to end the centrifugation run.
2.
Release the
Ramp-down takes place with the set brake level. The brake level is
displayed.
An audible signal sounds when the rotor comes to a standstill.
'OPEN' 'OEFFNEN' is displayed.
Personnel:
■
Trained user
[STOP/OPEN] button twice.
Press the
Ramp-down with brake level "9" (shortest ramp-down time) is dis-
played and executed.
The relative centrifugal force RCF is dependent on the speed and the centri-
fuging radius.
The relative centrifugal force RCF is stated as a multiple of the acceleration
due to gravity (g).
The relative centrifugal force RCF is a dimensionless numerical value and is
used to compare the separation and sedimentation performance.
Rev.: 13 / 01.2024
[STOP/OPEN] button lights up when the
[STOP/OPEN] button lights up when the rotor is
[START/IMPULS] button and the right side of the
[START/IMPULS] button.