5.3
Setting up and connecting the centrifuge
Setting up the centrifuge
AB2400en
6
5
4
Fig. 14: Transport lock
1
Bellows
2
Edge of the centrifuging chamber
3
Right side of device
4
Motor cover
5
Screws and spacer sleeves
6
Foam cylinder
2.
Remove 2 screws and 2 spacer sleeves (
3.
Keep the screws and spacer sleeves in a safe place.
4.
Open the lid.
5.
Remove the foam cylinder (
6.
Store the foam cylinder in a safe place.
7.
For MIKRO 200 R:
Check the bellows (
The edge of the centrifuging chamber (
the bellows.
WARNING
Risk of injury due to failing to maintain a sufficient distance to
the centrifuge.
−
As per EN / IEC 61010-2-020, no persons, hazardous
materials or objects may be present within a safety zone
of 300 mm around the centrifuge during a centrifugation
run.
−
A distance of 300 mm from the ventilation slots and ven-
tilation openings of the centrifuge must be maintained.
CAUTION
Risk of crushing and damage to the device due to it falling
down because of vibration-induced position alterations.
−
Place the device on a stable and level surface.
−
Select the installation surface dependent on the weight
of the device.
Rev.: 13 / 01.2024
3
5 ).
6 ).
1 ) under the motor cover ( 4 ) for correct seating:
2 ) must be in the groove of
Commissioning
1
2
19 / 39