Software operation
7.2.2
Opening or loading programs
7.2.3
Entering or changing programs
7.3
Cooling (for centrifuges with cooling)
7.3.1
Instructions, cooling
7.3.2
Standby cooling
28 / 39
[START/IMPULS] button.
4.
Press the
The setting is stored.
'*** lock ***' is displayed briefly.
If ON is set:
'*** ok ***' is displayed briefly.
If OFF is set:
1.
Use the
[SELECT] button to select the 'PROG RCL' parameter.
[Rotary knob] to set the desired program location.
2.
Use the
[START/IMPULS] button.
3.
Press the
'*** ok ***' is displayed briefly.
The centrifugation data of the desired program location is dis-
played
4.
To check the parameters: Press the
5.
To exit the parameter indicator: Press the
not press any button for 8 seconds.
1.
Retrieve program.
2.
If required: Press the
'> <' ).
indicator (
3.
If required: Press the
and set it with the
The parameters t/min and t/sec must be set to 0 with the
knob] to set continuous operation. Continuous operation is shown in
'¥' .
the indicator with
4.
Use the
[SELECT] button to select the 'PROG STO' parameter.
[Rotary knob] to set the desired program location.
5.
Use the
[START/IMPULS] button.
6.
Press the
Settings are stored in the desired program location.
'*** ok ***' is displayed briefly.
The settings are always stored in program location # if the
IMPULS] button is pressed without the 'PROG STO' parameter
being selected.
The temperature setpoint can be adjusted from -10 °C to +40 °C.
The lowest achievable temperature is rotor dependent.
After a centrifugation run, standby cooling is delayed and the display shows
'Lid unlocked' .
The delay time is adjustable from 1 to 5 minutes, in 1-minute increments. It
is preset to 1 minute.
■
The rotor is stationary.
■
The lid is open
Rev.: 13 / 01.2024
[SELECT] button several times.
[OPEN/STOP] button or do
[RCF] button to toggle between RPM and RCF
[SELECT] button to select the desired parameter
[Rotary knob] .
[Rotary
[START/