Télécharger Imprimer la page

Hettich MIKRO 200 R Mode D'emploi page 53

Masquer les pouces Voir aussi pour MIKRO 200 R:

Publicité

Les langues disponibles

Les langues disponibles

3
Device overview
3.1
Technical data
Manufacturer
Model
Type
Mains voltage (±10%)
Mains frequency
power consumption
Power consumption
max. capacity
max. permissible density
max. speed (RPM)
max. acceleration (RCF)
max. kinetic energy
Obligation to perform
checks
(DGUV Rules 100-500)
(valid only in Germany)
AB2400en
NOTICE
Damage to the device electronics due to incorrect voltage or
frequency at the device circuit breaker.
NOTICE
Damage to the device and samples due to premature pro-
gram termination.
Premature program termination is caused by power failure,
switching off during the program or pulling out the mains
plug.
Andreas Hettich GmbH & Co. KG,
D-78532 Tuttlingen
MIKRO 200
2400
200-240 V 1~
50-60 Hz
240 VA
1.2 A
30 x 2.0 ml
1.2 kg/dm³
15000
21382
5800 Nm
No
Rev.: 13 / 01.2024
Operate the device with the correct mains voltage and
mains frequency.
The value can be found in the technical data and on the
rating plate.
Do not switch off the device while the program is run-
ning.
Do not trigger the emergency release on the device
while the program is running.
Do not pull out the mains plug while the program is
running.
2400-01
110-127 V 1~
50-60 Hz
270 VA
2.7 A
Device overview
9 / 39

Publicité

loading

Ce manuel est également adapté pour:

Mikro 200