6.4.2
Lid with screw cap without hole
1
Fig. 17: Biosafety system
1
Rotary handle
2
Lid
3
Rotor
Closing
Opening
6.5
Centrifugation
6.5.1
Centrifugation in continuous operation
6.5.2
Centrifugation with time preselection
AB2400en
2
3
1.
Place the lid (
2.
Turn the lid (
closed.
1.
Turn the lid (
2.
Remove the lid (
Personnel:
■
Trained user
1.
Set minutes and seconds at
program.
[START/IMPULS] button.
2.
Press the
The centrifugation run is started.
[START/IMPULSE] button lights up during the centrifugation
The
run.
The timing starts at
The rotor speed or the RCF value, the temperature in the centri-
fuging chamber (only for centrifuges with cooling) and the elapsed
time are displayed during the centrifugation run.
[STOP/OPEN] button to cancel the centrifugation run.
3.
Press the
Ramp-down takes place with the set brake level. The brake level is
displayed.
An audible signal sounds when the rotor comes to a standstill.
'OPEN' 'OEFFNEN' is displayed.
Personnel:
■
Trained user
1.
Set centrifugation parameters or retrieve a program.
[START/IMPULS] button.
2.
Press the
The centrifugation run is started.
[START] button lights up during the centrifugation run.
The
The rotor speed or the RCF value, the temperature in the cen-
trifuging chamber (only for centrifuges with cooling) and the
remaining time are displayed during the centrifugation run.
Rev.: 13 / 01.2024
2 ) centrally on the rotor ( 3 ).
2 ) at the rotary handle ( 1 ) clockwise until it is tightly
2 ) at the rotary handle ( 1 ) anticlockwise until it is open.
2 ) from the rotor ( 3 ).
'¥' or retrieve a continuously running
'00:00' .
Operation
25 / 39