2. On the S S c c a a l l e e screen, press the A A d d d d E E q q u u i i p p . . button (E) ( Footboard control panel - Scale (page 34)).
3. One item at a time, add or remove the desired equipment from the product.
N N o o t t e e - - Do not touch the product while the product weighs the equipment.
4. Press the S S a a v v e e button to save the current weight and equipment count.
F F o o o o t t b b o o a a r r d d c c o o n n t t r r o o l l p p a a n n e e l l - - B B e e d d e e x x i i t t
The B B e e d d E E x x i i t t screen displays the bed exit functions of the product.
A
B
E
A
Home
B
Low
C
Medium
High
D
E
Turn off
A A r r m m i i n n g g o o r r d d i i s s a a r r m m i i n n g g b b e e d d e e x x i i t t
W W A A R R N N I I N N G G - - Do not use bed exit to replace patient monitoring protocol. Bed exit is intended only to aid in the detection of a
patient exiting the product.
When armed, bed exit monitors the patient's position on the product.
To arm bed exit:
1. Set the scale to zero/tare if not already performed. See Zeroing/taring the scale (page 35)
N N o o t t e e - - If you do not set the scale to zero before you place a patient on the product, bed exit may not operate as
intended.
2. Position the patient on the product.
3009-009-005 Rev AB.0
Returns to the H H o o m m e e screen ( Footboard control
panel - Home (page 32))
Allows the patient to move freely on the bed. Alarms
when the patient moves 50 percent body weight out
of the designated zone.
Allows for limited movement. Alarms when the
patient approaches the siderail or the foot end of the
bed.
Allows minimal movement. Alarms when the patient
moves out of the tightly restricted zone.
Turns off bed exit
C
D
37
EN