MAINTENANCE
Sunrise recommends that all fasteners be checked for wear, such as
loose bolts or broken components every 6 months. Loose fasteners
should be retightened according to the installation instructions. All
fasteners should be tightened to torque specifications as stated in
the hardware installation section. Contact your authorized Sunrise
Medical supplier immediately to replace any broken components.
Do not continue to use the system after identifying loose or broken
components.
EASY CLEANING
NOTE: When disassembling the back for cleaning, note the position
of the Velcro® Spine Align Components for correct reassembly.
To clean back cover
Remove cover by detaching Velcro. Detach cover from back shell.
Locate zipper on bottom of the cover and unzip. Remove foam and
turn cover inside out. Zip cover closed to avoid snagging. Machine
wash in warm water (60°C). Drip or tumble dry on low heat.
NOTE:
•
Do not dry clean covers or use industrial washers and
dryers to clean cover.
•
Do not steam autoclave.
•
Do not bleach.
Reinsert the foam into the cover. Please note, the base has two layers
of foam. When reinserting the foam, ensure that the thinner "light grey"
foam is facing the shell. Once the foam is secured inside the cover, zip
cover closed.
To clean back shell
Shell may be wiped clean with soap and water. No other
maintenance should be required.
To clean positioning components
Positioning components can be wiped clean with soap and water. Do
not immerse positioning components underwater.
If you have any additional questions regarding cleaning procedures,
please contact your authorized Sunrise Medical dealer.
Hygiene measures when being re-used
Prior to the wheelchair being re-used, it must be carefully prepared.
All surfaces which come into contact with the user must be treated
with a disinfection spray. To do this, you must use a disinfectant as
authorised/recommended in your country, for rapid alcohol-based
disinfection for medical products and medical devices, which must be
disinfected quickly. Please be aware of the manufacturer's instructions
for the disinfectant you are using. In general, a complete disinfection
cannot be guaranteed on seams. We therefore recommend that you
dispose of seat and back slings to avoid micro-bacterial contamination
with active agents according to your local infection protection law.
Conditions for safe storage
•
Products should be stored in a dry place: at 20 – 75% Relative
Humidity.
•
Protect from direct sunlight and dust for long periods.
•
Away from direct heat and at a temperature between 5°C (40°F)
to 40°C (104°F)
•
Position and stack the packaged products according to the "this
side up" arrows on the packaging.
•
If a package is high and narrow and therefore potentially unstable,
ensure it is well secured on the pallet or in the racking, to prevent
it from falling down.
•
Do not stack anything on top of the packaged products when the
"do not stack" mark on the packaging is shown.
•
Product should never be stored outside and subject to the
elements.
J3 Back 106109
M A I N T E N A N C E A N D C L E A N I N G
Laundry Care Symbols
7
Machine wash in warm
water (60° C)
Only non-chlorine bleach,
when needed
Drip dry
Do not iron
Do not dry clean
Rev.M