– Your multimeter can display diode forward bias of up to approximately 3.1 V. The
NOTE
forward bias of a typical diode is within the range of 0.3 V to 0.8 V; however, the reading
can vary depending on the resistance of other pathways between the probe tips.
– If the beeper is enabled during a diode test, the multimeter will beep briefly for a
normal junction and sound continuously for a shorted junction below 0.050 V. See
"Changing the beep frequency" on page
Reverse the probes (as shown in Figure 2-19) and measure the voltage across the diode
again. Assess the diode according to the following guidelines:
– A diode is considered good if the multimeter displays
– A diode is considered shorted if the multimeter displays approximately 0 V in both
– A diode is considered open if the multimeter displays
Figure 2-17
Keysight U1281A/U1282A User's Guide
forward and reverse bias modes, and the multimeter beeps continuously.
bias modes.
Open diode display
Making Measurements
129
to disable the beeper.
in reverse bias mode.
in both forward and reverse