Troubleshooting
Fault description
VERS. ERR 61.154
9.2
Performing a MAINS RESET
9.3
Emergency release
1
Fig. 20: Emergency release
1
Hole
9.4
Replacing the mains input fuse
40 / 44
Cause
Invalid machine version.
1.
Set the mains switch to
2.
Wait 10 seconds.
3.
Set the mains switch to
The lid cannot be unlocked by the motor in the event of a power failure.
Emergency unlocking by hand must be performed.
WARNING
Risk of electric shock due to maintenance and servicing
work on live device.
−
Disconnect the device from the mains before carrying
out repairs and maintenance.
WARNING
Danger of cutting and crushing due to moving rotor.
−
Do not open the lid until the rotor has stopped.
Personnel:
■
Trained user
1.
Look through the window in the lid to ensure that the rotor is sta-
tionary.
2.
Insert the hex key horizontally into the hole (
until the lid opens.
3.
Remove the hex key from the hole (
WARNING
Risk of electric shock due to maintenance and servicing
work on live device.
−
Disconnect the device from the mains before carrying
out repairs and maintenance.
Personnel:
■
Trained user
Rev.: 07 / 10.2023
Remedy
■
Perform a MAINS RESET.
[0] .
[I] .
1 ) and turn anticlockwise
1 ).
AB1101en