Safety
8 / 44
WARNING
Dangers due to insufficient maintenance or maintenance not
carried out on time.
−
Follow maintenance intervals.
−
Check the device for visible damage or defects.
If any visible damage or defects are present, take the
device out of service and inform a service technician.
WARNING
Risk of electric shock due to ingress of water or other liq-
uids.
−
Protect the device against external liquids.
−
Do not pour any liquids into the interior of the device.
−
Transport using original transport packaging.
WARNING
Contamination with hazardous substances and substance
mixtures!
Observe the following actions for substances and substance
mixtures that are toxic, radioactive and/or contaminated with
pathogenic microorganisms:
−
As a rule, use only centrifuge tubes with special screw
caps for hazardous substances.
−
Use sealable centrifuge tubes with a biosafety system
for materials of risk groups 3 and 4.
−
If no biosafety system is used, the device is not micro-
biologically tight in the sense of standard EN / IEC
61010-2-020.
−
Contact the manufacturer if necessary.
CAUTION
Risk of injury due to rotating rotor
Long hair and items of clothing can get caught on the rotor if
the rotor is moved manually.
−
Tie long hair back.
−
Do not allow garments to hang in the centrifuging
chamber.
NOTICE
Damage to the device electronics due to incorrect voltage or
frequency at the device circuit breaker.
−
Operate the device with the correct mains voltage and
mains frequency.
The value can be found in the technical data and on the
rating plate.