7.3
Rotor detection
7.4
Machine Menu
7.4.1
Querying system information
AB1101en
■
Rotor detection is performed after starting a centrifugation run.
■
The centrifugation run is cancelled after rotor detection if the rotor code
of the rotor is being read in for the first time. The rotor code (R) and the
newly detected rotor's maximum speed (maxRPM) permissible for the
centrifuge are displayed.
1.
Press any button after the rotor has stopped.
The maximum permissible number of run cycles (
played.
2.
Set the maximum permissible number of run cycles of the rotor or the
bucket.
■
Every rotor in a centrifuge has a maximum permissible speed.
The maximum rotor speed (
centrifugation run has started.
The centrifugation run is cancelled if the set speed is greater than
the permissible maximum rotor speed. The permissible maximum rotor
speed is displayed.
Adjust the speed to the maximum rotor speed.
■
If the cycle counter is enabled, the remaining number of run cycles (cen-
trifugation runs) of the rotor code used is displayed briefly after opening
the lid.
The following system information can be queried:
■
Centrifuge model
■
Maximum speeds of the various rotor codes
■
Centrifuge program version
■
Frequency converter type
■
Program version for the frequency inverter
The rotor is stationary.
1.
Press and hold the
'*MACHINE MENU*' is displayed after 8 seconds.
[SELECT] button.
2.
Press the
'-> Info' is displayed.
[START/PULSE] button.
3.
Press the
The centrifuge model is displayed.
[SELECT] button.
4.
Press the
The following is displayed:
'R' : Rotor code of the rotor
'*' : The rotor code of the rotor currently being used is marked with
an asterisk.
'RPMmax' : Maximum speed of the centrifuge rotor
5.
If necessary: press
The maximum speeds of the various rotor codes are displayed.
[SELECT] button.
6.
Press the
The centrifuge program version
Rev.: 07 / 10.2023
'ROTOR MAX' ) is displayed briefly after the
[SELECT] button.
[Adjustment buttons] .
'CP FW=' is displayed.
Software operation
'Cyc lim' ) is dis-