The precision is guaranteed for a reference junction temperature of 0° C.
When using the internal reference junction (except couple B) add an additional uncertainty of 0.2° C at 0° C. For further
temperatures, account must be taken of the sensitivity of the thermocouple to the temperature (T) in question, giving an
additional uncertainty of 0.2° C * S(0° C)/S(T).
Temperature coefficient: < 10 % of the accuracy/° C.
Display in ° C, ° F and K.
It is possible, except for the thermocouple B, to choose by programming the position of the cold junction with the
keyboard:
external at 0° C,
internal (compensation for the temperature of the terminals of the unit).
by programming the temperature.
D.3.5
Temperature by resistive probes
Type of sensors:
Pt 50ohm, 100ohm, 200ohm, 500ohm, 1,000ohm with α = 3851 as per CEI 751/1995
Pt 100ohm with α = 3916 as per JIS C 1604/1989
Pt 100ohm with α = 3926 as per EIT90
Ni 100ohm, 1,000ohm with α =618 as per DIN 43760
Ni 120ohm with α =672 as per MIL-T-24388
Cu 10ohm with α =427 as per MINCO 16/9
Cu 50ohm with α =428 as per OIML R 84
Sensor
Measuring range
Pt 50 (α = 3851)
- 220° C to + 850° C
- 220° C to + 850° C
Pt 100 (α = 3851)
- 200° C to + 510° C
Pt 100 (α = 3916)
- 210° C to + 850° C
Pt 100 (α = 3926)
- 220° C to + 850° C
Pt 200 (α = 3851)
Pt 500 (α = 3851)
- 220° C to + 850° C
- 220° C to + 850° C
Pt 1 000 (α = 3851)
- 60° C to + 180° C
Ni 100 (α = 618)
Ni 120 (α = 672)
- 40° C to + 205° C
- 60° C to + 180° C
Ni 1 000 (α = 618)
Cu 10 (α = 427)
- 70° C to + 150° C
- 50° C to + 150° C
Cu 50 (α = 428)
For negative temperatures, use the value displayed L and not its absolute value.
Temperature coefficient: < 10 % of the accuracy/° C.
The above accuracy is given for a 4-wire connection to the temperature measurer in continuous mode and for a 1 mA
measurement current for PT50, PT100, NI100, NI120, CU10 and CU50 probes and 0.1 mA for PT200, PT500 and PT1000
probes. For measurement currents included in the measurement range (0.1 mA-1 mA), the constant accuracy term for 1
year has to be multiplied by 2 (e.g., PT100 at 0.1 mA, accuracy is 0.006% + 0.07° C).
Taking into account, also, the intrinsic error of the temperature sensor used and its conditions of use.
"Transmission / Simulation" Function
Resolution
Accuracy / 1 year
0.01° C
0.006% R + 0.04° C
0.01° C
0.006% R + 0.035° C
0.01° C
0.006% R + 0.035° C
0.01° C
0.006% R + 0.035° C
0.01° C
0.006% R + 0.04° C
0.01° C
0.006% R + 0.04° C
0.01° C
0.006% R + 0.035° C
0.01° C
0.006% R + 0.04° C
0.01° C
0.006% R + 0.04° C
0.01° C
0.006% R + 0.04° C
0.01° C
0.006% R + 0.1° C
0.01° C
0.006% R + 0.05° C
D
NTA47271-300A2
181/184