CALYS 1500
To adjust the CALYS 1500, use the ADJUST function key. Make adjustments in the following order:
V1 measurement: run all calibration procedures
V2 measurement: run all calibration procedures
Ohms Offsets
Auto-Adjust
Transmission V2
For each type of calibration, select the function to be calibrated with the Up and Down navigation keys and follow the
instructions shown in the dialogue boxes.
When a unit is calibrated, data and certificate no. may be defined.
Date of calibration
To change the date of calibration and enter the reference of any calibration certificate, use the ADJUST
calibration menus.
A.5.3
Cleaning
112/184
NTA47271-300A2