To select a line from the list, use the Up and Down navigation keys.
To cancel a selected line, use the
The Format and Units zones are used to select the number of decimal places and the display units.
C.3 Differential measurements
The relative measurement function of the unit is used:
to program a reference value other than that of the unit (ZERO function),
to cancel by measurement or programming a constant or interfering value (TARE function).
When one of the relative measurement functions is active, the
window.
The ZERO
Define... menu is used to program the Tare value (positive or negative). This value is subtracted from the
measurements:
Value Displayed = Value measured – Tare Value
C.4 Calibrated sensors
key.
Calibrated sensors
symbol is displayed on the screen in the measurement
NTA47271-300A2