AOIP CALYS 1500 Mode D'emploi page 125

Calibrateur de process multifonction

Publicité

Les langues disponibles

Les langues disponibles

Enter device address from 00 to 15 or press F4 key (Scan) to scan for connected devices. The CALYS 1500 will then pool
the 15 network addresses to find out which replies and at which address.
For addresses 01 to 15, the unit is configured to fixed current. For address 00, the unit is configured to point-to-point mode
which uses the current loop over its full dynamic range and not as fixed current.
If only one device is available, enter 00 as "Device address" to use point-to-point mode. The sensor must also be
configured at address 0. Otherwise, scan to find out sensor address and change using "Configure" and "HART output"
menus (see details below).
Typical scan with only one device connected at address 00.
When selecting one of the addresses from a detected device, display is the same as if a valid address had been directly
entered:
Display depends on the type of sensor, in this case, it is a temperature sensor in point-to-point mode at address 0.
Pressing F1 key ("HART") displays a specific menu:
Measurement on channel V1
NTA47271-300A2
B
125/184