The following ranges are available:
Range
Resolution
Measurement current
Settling time
If the measurement current is very high, the message "Out of Range" is displayed in the transmission window.
The resistance simulation function can be used with either 2, 3 or 4 wire connection.
If a polling acquisition system is used, ensure that the current is maintained for more than 1 ms to avoid
measurement errors due to the response time of the resistance simulation function.
B.3.1.4
Resistive probe simulation (temperature)
Display the TRANSMISSION CONFIGURATION dialogue box:
Select the Rt transmission function, then the appropriate "type of probe", and range using the function and
navigation keys.
Select the display unit
Select measurement current (direct or pulsed)
Select current range
Confirm with ENTER.
Connection is made between the two
The following probes are available:
Sensor
Pt 50 (α = 3851)
Pt 100 (α = 3851)
Pt 100 (α = 3916)
Pt 100 (α = 3926)
Pt 200 (α = 3851)
Pt 500 (α = 3851)
Generation / Simulation or Measurement on Channel V2
400 Ohm – 1 mA
400 Ohm – 4 mA
1 mOhm
1 mOhm
0.1-1 mA
1-4 mA
< 10 ms
< 10 ms
terminals.
3600 Ohm
10 mOhm
0.1-1 mA
< 10 ms
Probe Type Caption
Pt 50
Pt 100
Pt 100-3916
Pt 100-3926
Pt 200
Pt 500
NTA47271-300A2
B
143/184