When set to the square law scale, the calibrator takes the square root of the input and displays the result as a percentage.
For example, if the calibrator is connected to the output of a differential pressure transmitter, it displays a result
proportional to the flow rate.
If the input current x varies between a and b, the scale is set according to the formula:
In the case of the 0-20 mA range, the scale curve is as follows:
In the case of the 4-20 mA range, the scale curve is as follows:
The CALYS 1500 displays in the window details of the selected configuration using the following icons:
: to show loop power supply off
: to show loop power supply on
: to show square law scale
Measurement on channel V1
) (
(