Télécharger Imprimer la page

AOIP CALYS 1500 Mode D'emploi page 175

Calibrateur de process multifonction

Publicité

Les langues disponibles

Les langues disponibles

D.2.1
DC Voltage
Range
Measuring range
+100 mV
- 10 mV to + 100 mV
+1 V
- 100 mV to + 1 V
+10 V
- 1 V to + 10 V
+50 V
- 5 V to + 50 V
Temperature coefficient <5 ppm/° C from 0° C to 18° C and from 28° C to 50° C.
Use the absolute value of the value measured (|L|) to calculate the accuracy.
D.2.2
DC Current
Range
Measuring range
0-20 mA
0 mA to 24 mA
4-20 mA
3 mA to 24 mA
100 mA
0 mA to 100 mA
Temperature coefficient < 7 ppm/° C from 0° C to 18° C and from 28° C to 50° C.
Loop power supply = 24 V typical (30 V max)
HART compatibility: input impedance Rin = 280
Linear or square law display scale.
These specifications are provided for the following configurations:-
- CALYS 1500 active mode (+24 V ON) 
- CALYS 1500 passive mode (+24 V OFF) 
Use the absolute value of the value measured (|L|) to calculate the accuracy.
D.2.3
Resistance
Range
Measuring range
400
0
to 400
3600
0
to 3600
50 K
(1) (2)
0
to 50 K
(1) Range on channel 1 only.
(2) See chapter D.1
Temperature coefficient < 5 ppm/° C from 0° C to 18° C and from 28° C to 50° C.
Automatic detection of connection scheme: 2 wire, 3 wire or 4 wire.
For 2 wire connection, the measurement includes the resistance of the line.
For 3 wire connection, add the out-of-balance of the line resistances.
Open circuit terminal voltage: < 10 V.
Continuity test:
Open circuit for R > 1000
Closed circuit for R < 1000
Measurement Function (Channel 1 and Channel 2)
Resolution
Accuracy / 1 year
1 µV
0.005% R + 2 µV
10 µV
0.005% R + 8 µV
100 µV
0.007% R + 80 µV
1 mV
0.007% R + 0.5 mV
Resolution
Accuracy / 1 year
0.1 µA
0.007% R + 0.8 µA
0.1 µA
0.007% R + 0.8 µA
0.1 µA
0.009% R + 2 µA
Measurer passive mode (+24 V OFF).
Measurer active mode (+24 V ON)
Resolution
Accuracy / 1 year
1 m
0.006% R + 8 m
10 m
0.006% R + 50 m
100 m
0.008% R + 1
D
Notes
Rin > 1 G
Rin > 1 G
Rin = 1 M
Rin = 1 M
Notes
Rin < 30
Rin < 30
Rin < 30
Notes
4-wire measurement
4-wire measurement
4-wire measurement
NTA47271-300A2
175/184