C.5 Calibration procedure
The CALYS 1500 is capabable of creating a calibration report from a pre-defined procedure.
The number of procedures that can be recorded depends on the size of the available memory and the size of each
procedure (number of test points). If the memory is not being used by other functions, it is possible to record several tens
of procedures.
To find the available memory size, refer to the chapter, Storage of current acquisitions (chapter C.6).
Open the MEASUREMENT MENU dialog box,
Select the Calibration procedure function,
Press ENTER.
Use the Up
/ Down
new procedure. A procedure may be duplicated with the F2 key (Duplicate). The F4 key is used to delete a
selected procedure.
Press ENTER.
With edit keys, enter "Device name" which is displayed in the "Reference" field in the above list.
Enter "Manufacturer" name which is displayed in the "Manufacturer" field in the above list.
The instrument features 2 calibration methods: "By comparison" or "Standard generator". Select from drop-down list.
In both methods, select "Device measurement": "Channel 1" or "Keyboard entry".