Anexo A: Informações sobre compatibilidade eletromagnética
GUIDANCE AND MANUFACTURER'S DECLARATION- ELECTROMAGNETIC IMMUNITY:
This device is intended for use in the electromagnetic environment specified below. The user of this device
should make sure it is used in such an environment.
Basic EMC
Immunity Test Levels
standard
HOME HEALTHCARE
ENVIRONMENT
Electrostatic
±8kV contact
Discharge (ESD)
±15kV air
IEC61000-4-2
Electrical fast
±2kV for power supply
transient/ burst
line
IEC61000-4-4
±1kV for input/output line
Surge
± 1 kV line(s) to
IEC61000-4-5
line(s)
± 2 kV line(s) to earth
Voltage
Voltage Dips:
dips, short
i) 100% reduction for 0.5
interruptions and
period,
voltage variations
ii) 100% reduction for 1
on power supply
period,
input lines
iii) 30% reduction for
IEC61000-4-11
25/30 period,
Voltage Interruptions:
100% reduction for
250/300 period
Power frequency
30 A/m
(50/60Hz)
magnetic field
IEC61000-4-8
Conducted RF
3 Vrms
IEC 61000-4-6
0,15 MHz – 80 MHz
6 Vrms in ISM and
amateur radio bands
between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
Voltar ao índice
20
All manuals and user guides at all-guides.com
Compliance Levels
±8kV contact
±15kV air
±2kV for power
supply line
±1kV for input/output
line
± 1 kV line(s) to
line(s)
230V (U
)
(1)
T
Voltage Dips:
i) 100% reduction for
0.5 period,
ii) 100% reduction for
1 period,
iii) 30% reduction for
25/30 period,
Voltage Interruptions:
100% reduction for
250/300 period
30 A/m
6Vrms
(4)
836002-5210 V3.4
Electromagnetic
Environment-Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of this
device requires continued operation
during power mains interruptions,
it is recommended that the device
be powered from an uninterruptible
power supply or a battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
O equipamento de comunicação
portátil e RF móvel não deve ser
usado próximo de qualquer parte
deste dispositivo, incluindo os
cabos, a uma distância inferior à
recomendada e calculada segundo
a equação aplicável à frequência do
transmissor.
www.stryker.com