Télécharger Imprimer la page

Stryker EOLE DC Powered Support Surface Manuel D'exploitation page 15

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 27
Problem
Loss of power
Low pressure alarm noises
Patient is bottoming out
Air Mattress is not secure
Air cells fail to inflate
www.stryker.com
All manuals and user guides at all-guides.com
Troubleshooting
Solution
Check if the plug is connected to mains.
1. Check if the CPR is sealed.
2. Check if the air cell is broken.
3. Check if the connection tube is tightly secured.
4. Check if there is any leakage on air cells.
Pressure setting might be inadequate for the patient. Adjust comfort range
1 to 2 levels higher and wait for a few minutes for best comfort.
1. Check if all the snap buttons or straps of mattress are all securely
fastened.
2. Check if the mattress is fixed to the bed frame by elastic straps.
Make sure the air hose is not kinked, cracked, or split. Verify that the power
switch is illuminated, signifying the control unit has power. Verify that the
air hoses are fully inserted with a positive connection.
836002-5210 V3.4
Return To Table of Contents
15

Publicité

loading

Ce manuel est également adapté pour:

2871