Télécharger Imprimer la page

Stryker EOLE DC Powered Support Surface Manuel D'exploitation page 14

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 27
The control unit housing, tubing, and mattress should be cleaned between patients.
To clean, use water and a clean cloth to wipe down the Control Unit, power cord, hoses, mattress top
cover, and bottom cover. Do not use abrasive cleaners on the mattress. Note: Blood and other body
fluids must be thoroughly cleaned from all surfaces before applying disinfectants.
Apply disinfectants to the external surfaces of the control unit, hoses and mattress top cover, and bot-
tom cover by wiping. Stryker recommends a chlorine-based solution with a concentration less than or
equal to 1000 ppm or 70% alcohol twice a week.
To wash the top cover of mattress by washing machine with normal process under the temperature
60°C in 45 minutes.
It is not recommended to disinfect the internal parts of the mattress on a regular basis, but only as
needed for particular instance, the air cell could be wiped with a cloth and disinfectants as recommend-
ed above.
Wipe down the mattress with a clean, dry cloth to remove any excess of disinfectant.
If other detergent or other cleaning agent is used, choose one that will not have adverse chemical ef-
fects on the surface of the plastic case of the control unit, mattress cover and any other component of
the device.
When cleaning or disinfecting the support surface, ensure that no liquid is allowed to seep into the zip-
per area and watershed cover barrier (underside); fluids allowed to come in contact with the zipper may
leak into the support surface.
Avoid dust and proximity to dusty areas.
All components should be air dried thoroughly before use.
The waterproof cap of power switch should be on the power switch.
Avoid using sharp tools on the waterproof cap over the power switch.
Please reply to your distributor if the cap is broken or missed off.
WARNING
Do not use phenolic based products for cleaning.
Do no dry the mattress in direct sunlight.
Return To Table of Contents
14
All manuals and user guides at all-guides.com
Cleaning and Disinfection
836002-5210 V3.4
www.stryker.com

Publicité

loading

Ce manuel est également adapté pour:

2871