Preventative maintenance should be performed annually, at a minimum. A preventative maintenance
program should be established for all Stryker Medical equipment. Preventative maintenance may need to be
performed more frequently based on the usage level of the product.
CHECKLIST
_______
Cover zipper opens and closes properly and has no visible damage.
_______
No tears, rips, holes, cracks, or other openings in the mattress cover.
_______
Check labels for legibility, proper adherence, and integrity.
_______
Support surface cover straps and snaps are intact and are not damaged.
_______
Straps properly secure the support surface assembly to the crib.
_______
Components have not degraded or come apart.
_______
Check main power cord and do not plug if there is an abrasion or excessive wear.
_______
Check airflow from the air hose.
_______
Check the air hose if there is kink or breaks
_______
Verify proper operation of the unit before returning it to service.
_______
Check the waterproof cap of power switch.