mode into sleep mode if it has not been used for
the amount of time indicated. Entering the sleep
mode closes any active LIS/HIS connection. Ac-
cept the settings by OK, or reject by Cancel.
Saving changes in personal settings
After all adjustments to personal settings have
been made choose Save.
Saving setting to profiles for further use
On the main menu screen choose Profiles. Choose
Save as profile, select an empty profile and name
it or select a profile you'd like to modify, give the
profile a new name if needed and accept by OK.
Measurement flow
In Measurement flow settings the operator can
adjust or choose laboratory/work routine oriented
settings like Operator and Patient IDs, printing,
LIS transfer and some test specific parameters
(see Picture 23).
These settings can be selected for temporary use
by choosing Save after making the changes.
For further use the settings need to be saved to a
Profile. For continuous use the settings should be
configured with Set-up wizard: Settings → Meas-
urement flow → Maintenance → Basic settings.
Operator ID
is an identification of the user.
Operator ID OFF: The instrument does not
request an Operator ID.
Operator ID ON: An Operator ID has to be
given before every sample measurement,
and the ID is coupled with the test result.
Operator ID ON + Propose previous: The in-
strument suggests a previous ID to be used,
but it can also be changed.
Patient ID
is an identification of the patient sample.
Patient ID OFF: The instrument does not re-
quest a Patient ID.
Patient ID ON: A Patient ID has to be given be-
fore every measurement, and the ID is coupled
with the test result.
Test parameters
Some test specific parameters can be changed.
Changes require giving the ADMIN password,
...
which is QRGOSET. Choose
/ Test parameters
and the corresponding Test. A list of parameters to
be configured will be shown.
Picture 23
Measurement flow menu
Quality control (QC) checking
These settings are for quality controls. Quality
control lots can be used to automate QC check-
ing. Enter and edit information of quality control
lots here. If POCT1-A2 is in use, information of a
new lot to be sent to LIS/HIS can be entered here,
but cannot be edited manually.
QC checking can be configured to give a warn-
ing only or it can be configured to disable patient
measurements if the quality control measurement
is outside the critical limits. A successful qual-
ity control measurement will re-enable patient
measurements. Setting QC checking OFF and ON
again will also reset QC lock and re-enable patient
measurements.
PROFILE
Settings
Operator ID:
Patient ID:
Test
parameters
...
OFF
OFF
ON
ON
QC
...
Propose
previous
Cancel
Maintenance
10:30
MEASUREMENT FLOW
Printing:
LIS transfer:
OFF
OFF
ON
ON
Automatic
Automatic
Save
USAGE
23